Applied Systems Engineering's Products Support from Applied Systems Engineering Downloads from Applied Systems Engineering Training Shop at Applied Systems Engineering Request a quote from Applied Systems Engineering Frequently Asked Questions of Applied Systems Engineering About Applied Systems Engineering Links important to Applied Systems Engineering
  Applied Systems Engineering - tab
Customer technical support
Customer technical support overview
ASE2000 Support
ASE2000 Knowledge Database
ASE2000 Release Notes
SPT4-NET Knowledge Database
Protocol PAK Knowledge Database
 
*

ASE2000 Support

Release Notes

The ASE2000 Support Pages has added a specific section for Release notes. This list describes what has changed from one version to the next. Please note the table is organized from most recent release at the top, to the oldest on the bottom.

Release Notes
Release
Details
Date Released
Version 1.52

Area

Protocol

Comment

New Features

Southern Services

Added support for Setpoint command

BCOM-USB I/O Driver

Added a signed driver for operation under Windows 7 64-bit mode

File Locations under Windows 7

Under Windows 7, user files are located in a user data area instead of the ASE2000 Program Files installation folder. Files created by operation of a previous ASE2000 version are copied from the old location to the new location on installation of release 1.52. Some files may not be accessible by an older ASE2000 version after 1.52 is installed.

Corrections

IEC 870-5-101 and 104

Corrected Count field in transmission of Directory Call command from 0 to 1

Telegyr 6500

Corrected multi-point reporting

Valmet (Tejas) series V 2-byte RTU

Added LRC and CRC security options

07-12-10
Version 1.50

Area

Protocol

Comment

New Features

Modbus

Added support for floating point values

Added options to modify the work orders and/or byte ordering for two-word register values (long and floating point)
IEC 870-5-104 Added support in the Point values view by specifying RTU ID 0
DNP3 Added ability to parse security messages in Monitor Mode

Corrections

DNP3

Corrected errors with UDP operation. This problem was introduced in 1.47 and did not exist in earlier releases.

10-12-09
Version 1.47

Area

Protocol

Comment

New Features

All

Added support for a BCOM-USB device

Corrections

All

Corrected error in line filtering. This version now supports the ability to display messages from/to a specific RTU only

IEC 101 Corrected a problem with SBO sequences in balanced mode if the DIR bit is set in messages from the controlled to controlling station. In prior versions, the FCB bit would not change state between the select and execute steps. All other cases had, and still do, work OK.
In RTU simulation mode, differentiate a response to a Delay Measurement message based on cause of transmission being Activate or Spontaneous

Getac LP

Corrected reporting ordering for DI points

Telegyr 6500

Corrected interpretation of analog values

SES-92

In RTU simulation mode, changed default return code for controls to indicate success

04-30-09
Version 1.46

Area

Protocol

Comment

New Features

DNP LAN/WAN

Allow use of ports other than 20000

DNP

Improved exchange matching logic to consider exchanges that differ only by qualifier information

IEC 870-5-101 and 103 Slave

Added option for Short Ack to replace No Data in response to Class 1 and Class 2 requests

IEC 870-5-104 Master

Allowed transmission of simulated point values in reverse direction

IEC 870-5-101/104 Slave

Limits number of points to those that can fit in one block

Network

Improved network card device detection to support more network cards

Rockwell 5011

Added “Group” to point values view

Binary and Generic Async

Allowed editing of exchange properties

Corrections

Amtrak

Improved exchange matching to allow for certain tag word bits to be set

CDC Type 2

Corrected parsing of responses with an odd number of analog points

Conitel 2025

Corrected parsing of Time Update message

Conitel 2100H

Increased resync period to 15 milliseconds (it was 15 bits)

DNP Master

Corrected problem with SBO relay commands where the outstation response used Data Link confirm services

Added parsing for re-transmitted Data Link frames

Getac 7020-LP

Corrected processing of inter-word gaps

Getac 7024-BCH RTU Sim

Corrected responses coded with an odd number of analogs

Corrected abort when response to a CSEL contained a RENDX instead of just REND

IEC 870-5-101/104 and DNP

Do not issue execute portion of a SBO sequence after negative response to select

IEC 870-5-104

Respond to TESTFR command even if STARTDT was not previously issued

Indactic 33/41

Corrected security calculation error for measurand requests sent in extended addressing mode

Line Analyzer

Data collection used to trigger on a change in carrier signals. Added trigger on change in input data value, for use with constant carrier lines

PG&E

Added support for 13-bit analogs

Telegyr 6500

Correct processing of responses with multiple pulse accumulator data blocks

Changed interpretation of pulse data from LSB to MSB

Added option to parse analog values with an inverted binary offset representation

TRW System 9 Version B

Corrected processing of multi-block data responses

01-07-08
Version 1.44

Area

Protocol

Comment

New Features

COM Port Version

Added support for a USB dongle

Corrections

IEC 870-5-101

Improved processing confirmation of file ASDU types

Corrected parse of Directory response

SpaBus

Added exchange for “Read Output Data”

Raw Data Display

Increased size of raw data viewing area to accommodate larger data blocks such as used for file transfer

Amtrak

Corrected processing of RR response for control actions

04-12-06
Version 1.43

Area

Protocol

Comment

New Features

All

Issue warning when a “master simulation” session is started that includes an enabled relay control operation. User must acknowledge before proceeding

Modbus

When a holding register exchange is configured to request digital input data, the digital points are identified by bit positions 0 to 15 within each holding register address

Corrections

All

Improve “stop on communication error” event processing. Prior releases may have stopped one message too late

IEC 870-5-101 and 104

Change incorrect identification of ASDU Type 7 from unknown to Bitstring

IEC 870-5-103

Correct display of “Measurand 2” values in the Point Values view

Identify “Invalid” bit when set in a time field

Tejas 5

Correct calculation of starting index for status points

TRW 9550

Fix parsing of responses longer than eight blocks

12-13-05
Version 1.42

Area

Protocol

Comment

New Features

Amtrak

Add new protocol

DNP3

Add capability to write multiple binary output points (object 10, variation 1)

DNP3 and IEC

Identify messages with no matching entry in the exchange list as “Unknown Exchange”. In the past, no exchange name was shown

Corrections

CDC Type 1

Correct display of CRC byte for Scan responses

Conitel 2100H

In the point values view, differentiate points in Status Group 1 from those in Status Group 2

DNP3

Correct RTU Simulation response to control output requests spanning more than one Data Link frame

IEC 870-5-101/104

Process User Data [Confirm] response to control request in balanced/Master Simulation mode

SC 1801 version 6

Correct processing of ‘L’ bit in SOE responses

SpaBus

Correct parsing of response messages

08-11-05
Version 1.41

Area

Protocol

Comment

New Features

DNP3

Added support for BCD (object 101)

Added capability to send binary event with relative, unsynchronized, time

Corrections

Capture Files

Corrected possible abort if a long capture file name was entered. Also corrects issue which occasionally required entering a full path for the file name (e.g., C:\temp\capture.cpt instead of just capture.cpt)

DNP3

Corrected data link and application confirm processing to binary input event messages with relative time (Object 2, variation 3)

Corrected occasional issue that would stop communication if a response was received from an DNP address other than as specified in the Exchange defaults menu

Corrects transmission of Octet String objects

IEC 870-5-103

Response to a General Command (ASDU type 20) has been corrected to be a Time Tagged Message (ASDU Type 1)

Telegyr 6500

Corrects display of RTU number in Point Values view

Telegyr 8979

  • Corrected processor utilization problem when “A” and “B” cables were interchanged.
  • Corrected processing of trip/close responses with exception blocks
05-18-05
Version 1.40

Area

Protocol

Comment

New Features

All

Removed support for ISA-bus BCOM2 card under Windows 2000 and XP

Changed color of Point Values view to be more readable in sunlight

DNP3

Recognizes Data Set Object Groups

IEC 870-5-103

Added entry of Generic Command data set information for multiple data sets

Rockwell 5020

New protocol

Corrections

All

Corrected points values display for point indices 256 and greater

No longer required to un-check “Capture Enabled” option before viewing a capture file

Correctly capture file usage when name in a newly loaded configuration file is different from the default

IEC 870-5-101

RTU Simulation Mode: Corrected balanced mode responses to general interrogation requests

IEC 870-5-103

Master Simulation Mode: Corrected Generic Command transmission when no data sets are defined

IEC 870-5-104

RTU Simulation mode: Corrected common address values contained in responses to control output sequences

Redac 70H

Corrected parsing of 24-bit pulse accumulators

S/Comm 11.1

Fixed CRC positioning for long messages

Telegyr 8979

Corrected parsing of ADC responses with six points

Enhanced parsing of multiple block messages

Wisp+

Corrected interpretation of Digital Report exchange

11-12-04
Version 1.39
Area Protocol Comment
New Features DNP 3 Adds support for 2-bit digital input points (objects 3 and 4)
Obtains Virtual Terminal data from an external file, allowing entry of non-printable characters
Adds ForceError option to prevent Nack response to Data Link Confirm messages received before a Reset Link
IEC 870-5-104  In Master Simulation mode, restarts message transmission sequencing whenever the controlled station reconnects. This allows STARTDT to be transmitted on the initial connection and each subsequent reconnection.  
Adds “Step Position with Date/Time” exchange
SPS
Add support for Southwestern Public Services protocol
Corrections Capture Files Corrects read error for some large capture files
Conitel 2100H Inverts point identification for pass select and analog scan responses
DNP3
Corrects RTU Simulation-mode abort on reception of an Assign Class message
IEC 870-5-101 and IEC 870-5-104 Corrects problem reading messages with fewer octets than expected
IEC 870-5-103 Corrects cause of transmission code used in response to a General Command message
Network Protocols Corrects reading a protocol message arriving in multiple IP packets
Rockwell 501x Ignores state of unused bits in a data response header
03-03-04
For more information about the products and services Applied Systems Engineering provides, select from the main menu on the top or please feel free to contact us directly if you have any questions or would like to discuss your requirements with an application engineer.

You may contact us in Campbell, CA by phone at 408-364-0500, fax to 408-364-0550, or via email at products@ase-systems.com.

Normal business hours are 9:00AM to 6:00PM Pacific Time, M - F.

Products | Support | Downloads | Training | FAQs | About Us
Links | Site Map | Search | Contact Us | Back to Home


Applied Systems Engineering, Inc.
1671 Dell Avenue, Suite 200
Campbell, California 95008
Phone: (408) 364-0500 • Fax: (408) 364-0550
Email: products@ase-systems.com

*